Fluke 726 Precision Multifunction Process Calibrator

Process calibration designed specifically for process industries

The Fluke 726 is a precise and powerful, yet easy-to-use, field calibrator. It features broad workload coverage, calibration power, and unsurpassed accuracy needed by process professionals. It measures and sources almost all process parameters to calibrate almost anything. Use it to test sensors and valves, and to test and calibrate transmitters. The 726 interprets results without the help of a calculator and stores measurement data for later analysis.

Measurement capabilities

This precision multifunction calibrator measures volts, mA, pressure, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters. The 726 also sources and simulates volts, mA, thermocouples, RTDs, frequency to calibrate transmitters. Measures and/or sources pressure using when paired with any of the 50 Fluke 750P series Pressure Modules to test and calibrate pressure transmitters and other devices. The 726 can also:

  • Source mA with simultaneous pressure measurement to conduct valve and I/P test
  • Conduct enhanced flowmeter testing with frequency totalizer and frequency pulse train source mode

Designed for ease of use

The 726 has built-in features that make it easier to use in nearly any measurement situation. From its ability to store frequently used test setups, to programming custom RTD probe calibration constants for enhanced temperature measurement, or the integrated voltage input protection design that improves reliability, the 726 is built to perform. The use of two separate channels even allows you to measure, source, or view process signals simultaneously. The backlit display provides better visibility in low-light conditions, and the standard 3-year warranty provides extra confidence that the 726 will be there when you need it to be.

You can also use the 726 to:

  • Capture the set, reset, and deadband of a pressure switch with the integrated switch test
  • Calculate the transmitter error % to interpret calibration results without the use of a calculator
  • Perform fast linearity tests with the auto-step and auto-ramp features
  • Power a transmitter during test using 24 V loop supply and simultaneous mA measurement

Specifications

Measurement Accuracy
Voltage DC30.000 V0.01% + 2 mV (upper display)
20.000 V0.01% + 2 mV (lower display)
90.00 mV0.01% + 10 μV
-10.00 mV to 75.00 mV0.025% + 1 count (via TC connector)
Current DC24.000 mA0.01%+ 2 mA
Resistance0.0 to 4000.0 Ω± 0.015%
Frequency2.0 to 1000.0 CPM0.05% + 0.1 cpm
1.0 to 1000.0 Hz0.05% + 0.1 Hz
1.00 to 15.00 kHz0.05% + 0.01 kHz
Sensitivity1 V peak-to-peak-minimum
Pressure

Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.

Source Accuracy
Voltage DC100.00 mV0.01% + 10 μV
20.000 V0.01% + 2 mV
-10.00 mV to 75.00 mV0.025% + 1 count (via TC connector)
Current DC24.000 mA (Source)0.01% + 2 μA
24.000 mA (Simulate)0.01% + 2 μA
ResistanceRange, accuracy (percent of reading + floor) and excitation current5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5 mA
5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3 mA
400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA
1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA
Frequency2.0 to 1000.0 CPM0.05%
1.0 to 1000.0 Hz0.05%
1.0 to 10.0 kHz0.25%
10.0 to 15.0 kHz0.5%
Waveform5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure and source accuracyCu101.8°C
NI-1200.15°C
PT-100 (385)0.45°C
PT-100 (392)0.35°C
PT-100 (391-JIS)0.35°C
PT-200 (385)0.95°C
PT-500 (385)0.55°C
PT-1000 (385)0.35°C
RTD Resolution0.01°C (0.01°F)
Cu10.01 C0.1°C (0.1°F)
J0.4°C
K0.5°C
T0.4°C
E0.4°C
R1.4°C
S1.4°C
B1.4°C
L0.45°C
U0.45°C
N0.6°C
XK0.4°C
BP1.1°C
TC ResolutionJ, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
Specifications
Ramp functionsSource functionsVoltage, current, resistance, frequency, temperature
RampsSlow ramp, Fast ramp, 25% step-ramp
Loop power functionVoltage24 V
Accuracy10%
Maximum current22 mA, short circuit protected
Step functionsSource functionsVoltage, current, resistance, frequency, temperature
Steps25% of range, 100% of range
Environmental Specifications
Operating temperature-10°C to +55°C
Storage temperature-20°C to +71°C
Operating altitude3000 m
Safety Specifications
Agency approvalsEN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Mechanical and General Specifications
Size200 x 96 x 47 mm (7.9 x 3.8 x 1.9 in)
Weight650 g (23 oz)
Batteries4 AA alkaline batteries
Battery replacementSeparate battery compartment, accessible without breaking calibration seal
Side port connectionsPressure module connector, also used for remote real-time programming

Features

  • Precision multifunction process calibrator designed specifically for the process industry
  • Delivers precise measurement and calibration source performance with accuracies of 0.01%
  • Stores up to eight calibration results in memory for later analysis
  • Offers HART mode that inserts a 250 ohm resistor in mA measure and source for compatibility with HART instrumentation
  • Transmitter error % calculation, interpret calibration results without a calculator
  • Frequency totalizer and frequency pulse train source mode for enhanced flowmeter testing
  • Integrated pressure switch test allows you to capture the set, reset and deadband of a switch
  • Custom RTD curves, add calibration constants for certified RTD probes for enhanced temperature measurement
  • New voltage input protection design for improved reliability
  • Two separate channel to measure, source and view process signals simultaneously
  • Measure volts, mA, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters
  • Source/simulate volts, mA, thermocouples, RTDs, frequency, and pressure to calibrate transmitters
  • Source mA with simultaneous pressure measurement to conduct valve and I/P tests
  • Perform fast linearity tests with auto step and auto ramp features
  • Power transmitters during test using 24 V loop supply and simultaneous mA measurement
  • Store frequently-used test setups for later use
  • A backlit display provides better visibility in poor light
  • Features frequency totalizer and frequency pulse train source mode for enhanced flowmeter testing
  • Provides two separate channels to measure, source and view process signals simultaneously